Comprehensive drug lighting stability test chamber (UV lamp supervise and control)
◆Lighting system of the comprehensive drug lighting stability test chamber
●Comprehensive drug lighting stability test chamber accord with the requirement of lighting stability and relevant international manufacturing standards. Supported by the 2015 Pharmacopoeia drug stability test guidelines. Simulate a stable temperature, humidity and lighting condition of long time operation in a scientific way.
●Rack illumination system or outdoor illumination system for chosen. Visible modulator tube and UV lamp modulator tube included and independent controlled.
Comprehensive drug lighting stability test chamber (UV lamp supervise and control) Loading bracket adjustable.
◆Irradiance supervise and control
●Breakthrough the defect of irradiance can not be displayed and supervised, decrease the irradiance attenuation problem and error caused by the experiment. Provide the lighting sensor and UV lamp detector, convenient for the users to observe and calibrate.
◆Professional UV lamp modulator tube
●Compare to the normal UV lamp modulator tube, our professional UV lamp have the features of high quality and even spectral powder distribution, avoid the affection caused by the aging problem.
●Apply the UV lamp holder which can run normally under the high wet condition.
◆Humanized design
●Newly fluoride-free design, high efficiency, energy saving.
●Micro-computer controller: accuracy control, special sterilization 304 stainless steel liner, semicircle design(Four corners), easy cleaning.
●Special air circulatory system, ensure the temperature and humidity even inside.
●A φ25mm testing hole on the left side of the chamber
◆Continuous operation qualified
●2 sets imported compressor automatically switch, prevent the long time running defect from happening.
●Non-defrosting, continuous operation work normally.
◆Quality qualified
●Imported temperature and humidity controller, compressor and circulating fan, make the continuous operation possible.
◆Imported humidity sensor
●Apply the humidity sensor work under the high temperature.
◆Programmable touch screen controller(GSP series)
●Large LCD touch screen, easy operation and programmed.
●Language switch on the operation menu, data display.
●100 programs, 999 circular steps. Timing setting from 1~5999 min.
●After data and experiment parameters input, password lock screen runs, security control.
●P.I.D auto tuning function, amend the temperature and humidity change, accuracy control.
●RS-232 or RS-485 communication screen; Input process on computer, supervise the experiment process.
◆Data record and error display
●LCD screen will display the error information when the device breaking down.
●Printer or RS485 port inside; Temperature recording(Optional choice).
◆Wireless alarm system(SMS alarm system)-optional
●Message will be delivered to the users when error occurs at the first moment. Ensure the security.
◆Security function
●Independent temperature warning system; sound-light alarm.
●Too high/too low and over temperature also trigger the alarm.
●Standard with door Lock system, prevent the UV hurting people when the chamber under running.
●Password Lock Screen, security control.
※Provide 3Q verification and calibration services(Optional choice):
※Provide the IQ/OQ/PQ service to customers, meet the drug stability test chamber performance and requirement. Guarantee the highly accuracy of temperature and humidity data.(Optional)
※Provide the authority third party testing report in shanghai.(Optional choice)
Comprehensive drug lighting stability test chamber (UV lamp supervise and control) Execution and satisfying criteria:
Supported by the 2015 Pharmacopoeia drug stability test guidelines and GB/T10586-2006 relevant terms.
●Speed up experiment: 40℃ ±2.0℃/75%RH±5%RH; or 30℃ ±1.0℃/60%RH±5%RH 180 days.
●Long-term experiment: 25℃ ±2.0℃/60%RH±5%RH,or 30℃±2.0℃/60%RH±5%RH 365 days
●Highlight illumination experiment: 4500±500LX 10 days
●Meeting the Q1B illumination requirement: total illumination≥1.2×106LUX.hr;
Ultraviolet energy ≥200W.hr/m 2.
●Illumination and Ultraviolet irradiation test can be done together.
Comprehensive drug lighting stability test chamber (UV lamp supervise and control) Technical parameters:
Model | SM-LHH-150GSD-UV | SM-LHH-250GSD-UV | SM-LHH-500GSD-UV | SM-LHH-800GSD-UV | SM-LHH-1000GSD-UV | SM-LHH-150GP-UV |
SM-LHH-150GSP-UV | SM-LHH-250GSP-UV | SM-LHH-500GSP-UV | SM-LHH-800GSP-UV | SM-LHH-1000GSP-UV | SM-LHH-250GP-UV | |
SM-LHH-1500GSP-UV | SM-LHH-400GP-UV | |||||
Temperature control range | Without illumination: 0~65℃; with illumination: 10~50℃ | |||||
Temperature fluctuation | ±0.5℃ /±2℃ | |||||
Humidity control range/Humidity deviation |
35~95%RH/±3%RH | None | ||||
Illumination intensity | Cool white fluorescent 0-6000Lux adjustable/±500Lux, UV light intensity adjustable | |||||
Ultraviolet spectral range | 320~400nm | |||||
conform to test standard | ICH Q1B command, Cool white fluorescent≥1.2×106Lux.hr; ultraviolet fluorescence≥200 w.hr/m2 | |||||
Illuminant | Cool white fluorescent(conform to ISO 10977); UVA ultraviolet fluorescence | |||||
Standard light source | Cool white light(D65, 6500k)/ UV wavelength coverage:320-400nm | |||||
Refrigerating system | 2 sets imported compressor automatically switch | |||||
Controller | GSD series: programmable LCD controller; GSP series: programmable touch screen controller |
Programmabe LCD controller |
||||
Power supply | AC220V 50HZ | AC380V 50HZ | AC220V 50HZ | |||
Volume | 150L | 250L | 500L | 800L | 1000L/1500L | 150L/250L/400L |
Liner dimension | 550×405×670 | 600×500×830 | 670×725×1020 | 800×590×1650 | 1050×590×1650 | 550×405×670 |
(W×D×H mm) | ||||||
1550×590×1650 | 600×500×830 | |||||
700×550×1140 | ||||||
Outside dimension | 690×805×1530 | 740×890×1680 | 850×1100×1930 | 1360×890×2000 | 1610×890×2000 | 690×805×1530 |
(W×D×H mm) | ||||||
2110×890×2000 | 740×890×1680 | |||||
950×850×1850 | ||||||
Light way | Illumination on top/ Illumination on door | Illumination on top |
Illumination on door / Illumination on top |
|||
Embedded printer | standard configuration | |||||
Security system | Over temperature protection, water shortage protection, wire alarm system | |||||
Remarks | 1.High accuracy digital recorder | |||||
2.Manual stepless adjusted, standard with illumination detector and UV detector, UV illumination on top | ||||||
3.Multi-layer illuminator |