Best Suitable for sorting machine such as G, J, K, and M class, chip, melf, lead type resistor of sorting machine, taping machine, and coating conveyor
● Fast measurement (FAST):about 2.1msec. [60Hz]/2.5msec. [50Hz]
● Measuring range:0~999mΩ and 0.01Ω~99.9MΩ
● Display shows percentage indication by digital setting for a standard resistance value
● HI/GO/LO output installing with digital comparator
● Contact-check function (option) is selectable from premeasurement/postmeasurement/OFF
● Shift output and solenoid power source output (option)
● Measuring current/voltage-check are built-in as standard function.
Specifications
Measuring range | Measuring current | Accuracy | Indication range | |
---|---|---|---|---|
SLOW | FAST | |||
0.1Ω~0.999Ω | 100mA | within ±α±0.2%±1digit | within ±β±0.2%± (2+n) digit | ±99.9% |
1Ω~9.99Ω | 50mA | within ±0.2%±1digit | within ±0.2%±(2+n)digit | |
10Ω~99.9Ω | 5mA | |||
100Ω~999Ω | ||||
1kΩ~9.99kΩ | 0.5mA | |||
10kΩ~99.9kΩ | 50μA | |||
100kΩ~999kΩ | 5μA | |||
1MΩ~9.99MΩ | 0.5μA | within ±0.3%± (2+n) digit | ||
10MΩ~99.9MΩ | 0.05μA | within ±0.2%±2digit | ---- | ---- |
mΩ CHECK (0~999mΩ) | 50mA | within ±0.2%±1digit | within ±0.2%±2digit | 0~999mΩ |
※α:(500/setup value mΩ)×0.01%, β: (500/ setup value m Ω)×0.02%, n: when measuring time is F1, n=1, when measuring time is F2, n=0)
Measuring time | Power source frequency | Remote start | Free running | ||
---|---|---|---|---|---|
SLOW | FAST-1 (F1) | FAST-2 (F2) | SLOW | FAST | |
60Hz | about 16.7msec. | about 2.1msec. | about 4.2msec. | 30 times per second | 30 times per second |
50Hz | about 20.0msec. | about 2.5msec. | about 5msec. | 25 times per second | 25 times per second |
※Judgment time (from input start signal to output EOC signal) adds about 1.5msecond for the measuring time above (slightly differ from a range)
※When before contact check, it adds about 3.5msecond, and adds about 0.7msecond after contact-check for the judgment time above.
Open-circuit voltage of measuring terminal | less than 15V |
---|---|
Measuring method | 2-terminal measurement (more than 100kΩ) 4-terminal measurement (less than 99.9kΩ) |
Setting range for measurement value | ±00.0%~±99.9% both for high and low limit In a case of mΩCHECK:000mΩ~999mΩ |
Operation condition | [Temp.] 0℃~+50℃ [Humidity] less than 80% |
Power supply | AC100V/117V/220V/240V switchable, 50/60Hz, about 50VA |
Outer dimension | about 333 (W) ×85 (H) ×245 (D) mm (excluding protruding parts such as rubber legs, etc.) |
Weight | about 2.8kg |
The Outline
AX-150A can measure a wide range of resistance from 1mΩ to 99.9MΩ,ultra fast speed, high accuracy. The unit can be shown the measured value as a digital indication, and also outputs a signal outside to judge HI/GO/LO. It can be switched a measuring speed, selectable from 2 types of FAST, a type of SLOW.
The unit can be equipped Contact-check function as an option: When measures 4 terminals, if either one occurs contact failure, judges a detect regardless of its measured value, and outputs failure signal outside and C.E. (Contact-check error) signal all together.
The behavior of Contact-check is selectable from premeasurement/postmeasurement/OFF with a switch of a behind panel.
Also it sets [a checking circuit of an abnormal measuring current] has always watching a measurement current, outputs NG judgment when terminal trouble of contact condition occurred during measurement.
Either shift output-solenoid power supply or BCD date can be placed as an option.