Compare with external standard thermistor/polymer PTC element of resistance value Best suitable for percentage measurement
● Measuring range:
[% Measurement] -99.99%~+50.00%
[Absolute measurements] 0.00mΩ~15.000MΩ
● Compares with external standard thermistor or polymer PTC element, can be percentage measured for standard value regardless of temperatures characteristic at room temperature
● Deter from self-heating of measured object depends on applying measurement voltage for pulsating.
● Can be measured resistance value of external standard and measured object
● Can be corrected an error of external standard resistance value
● Installed digital comparator [up to 3 or 7 classification]
● Measuring current/voltage-check are built-in as standard function.
● Measurement current, a checking circuit of an abnormal measuring current as standard equipment
● Simplify the check for the each setting item by LCD display adoption
● RS-232C interface is equipped as a standard
Specifications
Measuring range and Accuracy (at23℃±5℃)
Range | Measuring range | Measuring current | Accuracy |
---|---|---|---|
10Ω | 0.00mΩ~15.000Ω | 316mA | within ±0.02%±α±1d |
100Ω | 0.000Ω~150.00Ω | 1.00mA | within ±0.02%±1d |
1kΩ | 0.000Ω~1.5000kΩ | 316μA | |
10kΩ | 0.00Ω~150.00kΩ | 100μA | |
100kΩ | 0.0Ω~1500.0kΩ | 20μA | |
1MΩ | 0Ω~1.5000MΩ | 3.16μA | |
10MΩ | 0Ω~15.000MΩ | 0.63μA | within ±0.03%±α±1d |
% | 1mΩ~15.000MΩ/±50.0% | refer to the above | refer to the above |
Compare with external standard thermistor/polymer PTC element of resistance value, Best suitable for percentage measurement
Open-circuit voltage of measuring terminal | [Slow]:AC1~10cycles, [Fast]:0.1mS~99.9mS |
---|---|
Measuring method | 4-terminal measurement (Available to contact check) |
Sampling time | [Remote start mode] about 1m [Fastest time] |
Comparator set range | [Resistance measurement]:0~15,000 counts both for HI and LO |
[% Measurement]:-99.99%~+50.00% both for HI and LO [Standard 100~10000:±50.00%] |
|
Indication of comparator’s comparison result | Comparator judgment result indication LO/LLG/LG/GO/HG/HHG/HI, judgment LED indication, buzzer setup |
Machine interface control signal (Connector:57-40240 Equivalent) |
Remote start input:”L” [0V] →”H” [DC12V] start |
Remote hold input:Open and “H” [DC12V] :Free run/”L” [0V]:Hold | |
comparison output [7 common signal classification]: open collector:max.80V, 100mA |
|
contact error output [CONT-E]:open collector:max.80V, 100mA | |
end of comparison output [EOC]:open collector:max.80V, 100mA | |
Operation condition | [Temp.] +5℃~+40℃ [Humidity] less than 85% (Disabled when condensation) |
Power supply | AC100V/117V/220V/240V selectable, 50/60Hz, about 60VA |
Outer dimension | 333 (W) ×99 (H) ×300 (D) mm (excluding protruding parts such as rubber legs, etc.) |
Mass | about 4kg |
The Outline
AX-1137A/B indicates percentage deviations compare resistance value of thermistor polymer PTC element with external standard thermistor polymer PTC element.
Adding comparator of 2 layers (1137A) 5 layers (1137B) outputs to judge up to 7 classifications compare with the setup value.
Either GP-IB or centronics output can be equipped as an option.