Use temperature | 50~360℃ | |
---|---|---|
Test Temperature, T | Maximum allowable deviation from required test temperature | |
At (10±1) mm above the top surface of the standard die | From (10±1) mm to (70±1) mm above the top surface of the standard die | |
125 ≦ T <250 | ±1℃ | ±2℃ |
250 ≦T<300 | ±1℃ | ±2.5℃ |
300 ≦T | ±1℃ | ±3℃ |
Counter accuracy | ±0.01s | |
Piston push rod weight | 100g | |
Automatic cut | Interval time 1-300 seconds can be set | |
Heater size | 9.55±0.01mm | |
Hole mold | 2.095±0.005mm | |
Counter weight | Stainless steel 1100g、2060g、3700g、4900g、9900g、12400g、19900g、21500g(Optional) (rod weight 100g) | |
Accessories | Cleaning brush、measuring cups、funnel and Die calibrationplug gauge | |
Weight | 30 kg (40kg with MVR part) | |
Dimension | 47x38x48cm | |
Optional | High MI value material blockmechanism, MVR test mechanism. |
?
- Cometech reserves ultimate modification right of product specification.