Model: AX-3245A.5 classification digital capacitance checker
Best suitable for inspection of aging process of aluminum electrolytic capacitor
● Contact check function is standard equipment
● Measurement frequency: 120Hz
● Measurement time: general measurement time/about 105msec.
(Contact& residual voltage check time about 27msec, + stabilizing time 45msec.)
At the time of free running about 3 times/second
● Installing with digital comparator, 5 classification judgment indication・output
Specifications
Measuring range and Accuracy (Series equivalent circuit D<1 at23℃±5℃)
Range | 400nF | 4μF | 40μF | 400μF | 4mF | Display | ||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
C Accuracy | ±0.5%±2 (digit) | 1%±5 (digit) | ||||||||||||||||||
tanδ Accuracy |
|
|
||||||||||||||||||
Measuring mode | serial equivalnet circuit |
※When tanδis 100.0%~199.9%, the accuracy of C or tan δ will be twice the above
Measuring mode | serial equivalnet circuit |
---|---|
Measuring signal level | less than 400mV (r.m.s.) |
Measuring method | signal application 2 terminals signal detection 2 terminals, five terminals measurement structure from guard terminal |
Measuring frequency | within 120Hz±0.1% |
Measuring terminal | Front panel section:5 pin terminal |
The rear panel:Metal Circular Connectors 8P | |
Contact check | The function to perform measurement error if contact resistance detects over 200Ωcontact check input terminal more than one terminal |
Input protection | The function to protect measurement circuit if non-electric discharge voltage of measurement sample detects over 2.2V on 4 terminals connection (stay voltage check) |
Measuring time | [HOLD mode measurement time] about 105msec. |
(Contact check & Check residual voltage) about 27msec. | |
(Time measurement stability) about 45msec. | |
(To open the measurement terminals) about 96msec. | |
[Measurement time free-running mode] about 3 times/sec. | |
Comparator setting | Capacity C:4 points LL, L, H, HH “200~3999” |
Loss tanδ:”1~199.9″ | |
Comparator decision | Digital data comparison method |
Judgment result display | Capacity:CLO, CLG, CGO, CHG, CHI |
tanδ:DGO, DNG | |
Total:C/D-GO | |
External output result of the judgment | GO1, GO2, CNG, DNG |
Control signal | Input:START, HOLD, RESET, CINH, SHIFT, NINH, GINH |
Output:GO1, GO2, CNG, DNG, CER, COV, DER, DOV, CEND MEND, BUSY, SLCT, CCE, VCE, Various shift determination |
|
Operation condition | [Temp.] 0℃~+40℃ [Humidity] less than 80% (No condensation) |
Power supply | AC100V/117V/220V/240V±10%, 50/60Hz, about 45VA |
Outer dimension | 432 (W) ×149 (H) ×450 (D) mm (excluding protruding parts such as rubber legs, etc.) |
Weight | about 12kg |