Máy kiểm tra điện dung / trở kháng kỹ thuật số AX-365E Adexaile ADEX
Ultra High speed and High Accuracy,
120Hz/1kHz Digital MLCC Checker
Improvement in Accuracy and Stability of Very Small Capacity Measuring by Adding 2pF Range.
● Available to change of series equivalent circuit and Parallel equivalent circuit.
● Available to measure the signal of constant voltage. (impossible for a part of range)
● Available to measure Tanδ. (impossible for 2pF range)
● Supplying measured current in order to reduce the function of Probe-Contact. [120μF/1.2mF range]
● Available to select 3 terminal/5 terminal (Contact Check) On each Range.
● Wide measuring range from very small capacity to large capacity. [2pF~1.2mF range]
● 3½ digit display and LO/GO/HI/Contact Error Output.
● Ultra High Speed measurement: Measurement time Typical value [1.2m sec: FAST mode]
● RS-232C interface as standard.
Specifications
Measuring range and Accuracy (D≦0.1 at23℃±5℃)
| Test Current | Measuring range | Resolution | Accuracy | Measuring voltage | |
|---|---|---|---|---|---|
| 2pF | 0.000pF~1.999pF | 0.001pF | within ±1.0% of rdg ±5digit×α |
1kHz, 1V±5% [rms] | 1kHz, 5V±5% [rms] |
| 20pF | 0.00pF~19.99pF | 0.01pF | within ±0.25% of rdg ±3digit×α |
||
| 200pF | 0.0pF~199.9pF | 0.1pF | within ±0.2% of rdg ±2digit×α |
||
| 2nF | 0.000nF~1.999nF | 0.001nF | within ±0.2% of rdg ±2digit |
||
| 20nF | 0 .00nF~19.99nF | 0.01nF | |||
| 200nF | 0.0nF~199.9nF | 0.1nF | |||
| 2μF | 0.000μF~1.999μF | 0.001μF | 120Hz, 0.5V±5% [rms] | ||
| 20μF | 0.00μF~19.99μF | 0.01μF | within ±0.3% of rdg ±3digit×α |
||
| 120μF | 0.0μF~120.0μF | 0.1μF | within ±1.0% of rdg ±5digit×α |
1kHz, 1V±5%~-40% [rms] |
|
| 200μF | 0.0μF~199.9μF | 0.1μF | within ±0.5% of rdg ±3digit×α |
120Hz, 0.5V±5% [rms] | |
| 1.2mF | 0.000mF~1.200mF | 0.001mF | within ±1.5% of rdg ±5digit×α |
120Hz, 0.5V±5%~-50% [rms] |
※in case of 0.1<D<1, 25D/100 (%) is added on the above accuracy.
α:2 at FAST (α=10 at 2.00pF 1.0V) 1 at SLOW (α=5 at 2.00pF 1.0V)
| Measuring method | 5-terminal measurement (Available to select the measuring method on each range), Parallel equivalent circuit. |
|---|---|
| Measuring frequency | 120Hz/1kHz±0.1%,SINE wave |
| Test signal output impedance | about 1.5Ω |
| Stray Capacity correcting range | about 20pF |
| Full scale and zero temp.coefficiecy | within ±100ppm/30℃ |
| Measuring time | [Free running] FAST:about 1~5 times/sec. SLOW: FAST×N (N:Setting number of average value) [The time from remote satart to INDEX signal] FAST:1.2m sec (typical value) |
| Comparator set range | Capacitance:1999 both for HI and LO [120μF, 1.2mF range: 1199 both HI and LO] Tan δ:99.9% |
| Operation condition | [Temp.] 0℃~+50℃ [Humidity] less than 85% |
| Power supply | AC85V~265V, 50/60Hz, about 50VA |
| Outer dimension | 333 (W) ×99 (H) ×300 (D) mm [excluding protruding parts such as rubber legs, etc.] |
| Weight | about 4.0Kg |
Cần thêm thông tin hoặc tư vấn kỹ thuật?
Nếu Quý khách cần thêm thông tin chi tiết về sản phẩm, thông số kỹ thuật, ứng dụng thực tế hoặc yêu cầu mẫu thử, đội ngũ của chúng tôi luôn sẵn sàng hỗ trợ.
📞 Điện thoại: (024) 6687 2330
📧 Email: sale@pemax-mte.com
Hoặc gửi yêu cầu qua biểu mẫu liên hệ hoặc trao đổi trực tiếp qua hộp chat ở góc phải màn hình.