Model: AX-369B 120Hz/1kHz High Speed Type Digital C-tanδ Meter
For the sorting meter of Aluminum Electrolytic Condensor and Tantalum Condensor
● Measuring frequency:120Hz/1kHz
● Measurement speed:23msec.
● Measuring range:C 0~39.99mF/tanδ 0~199.9%
● GP-IB, RS-232C, Printer output (8 bit parallel Centronics)[Option] (Only one kind of interface can be provided with the checker.)
● Connecting cable to printer [Option]
Specifications
Measuring range and Accuracy (at23℃±5℃)
Range | 1kHz | 40nF | 400nF | 4μF | 40μF | 400μF | 4mF | |||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
120Hz | 400nF | 4μF | 40μF | 400μF | 4mF | 40mF | ||||||||||||||
C Accuracy | ±0.3%±2 (digit) | ±γ%±2 (digit) |
±1%±5 (digit) | |||||||||||||||||
tanδ Accuracy |
|
|
||||||||||||||||||
Measuring mode | serial equivalnet circuit |
※When tanδis 100.0%~199.9%, the accuracy of C or tan δ will be doubled the above | β | : | 5 (120Hz)/10 (1kHz) |
γ | : | 1 (120Hz)/0.5 (1kHz) |
Measuring mode | serial equivalnet circuit |
---|---|
Measuring signal level | less than 400mV (r.m.s.) |
Measuring terminal | 5 terminals structure from voltage, current, and guard terminal |
Measuring frequency | 120Hz/1kHz±0.1%, sine wave form |
Measuring time | [Remote start mode] about 23msec. |
[Free running mode] about 3 times/sec. | |
DC bias | DC1.5V (built-in) and 0~25V (remote) |
Operation condition | [Temp.] 0℃~+50℃ [Humidity] less than 80% |
Power supply | AC100V/117V/220V/240V selectable, 50/60Hz, about 30VA |
Outer dimension | 333 (W) ×99 (H) ×300 (D) mm (excluding protruding parts such as rubber legs, etc.) |
Weight | about 4.7kg |