Máy Kiểm tra điện trở kỹ thuật số AX-1137B Adexaile ADEX
Compare with external standard thermistor/polymer PTC element of resistance value Best suitable for percentage measurement
● Measuring range:
[% Measurement] -99.99%~+50.00%
[Absolute measurements] 0.00mΩ~15.000MΩ
● Compares with external standard thermistor or polymer PTC element, can be percentage measured for standard value regardless of temperatures characteristic at room temperature
● Deter from self-heating of measured object depends on applying measurement voltage for pulsating.
● Can be measured resistance value of external standard and measured object
● Can be corrected an error of external standard resistance value
● Installed digital comparator [up to 3 or 7 classification]
● Measuring current/voltage-check are built-in as standard function.
● Measurement current, a checking circuit of an abnormal measuring current as standard equipment
● Simplify the check for the each setting item by LCD display adoption
● RS-232C interface is equipped as a standard
Specifications
Measuring range and Accuracy (at23℃±5℃)
| Range | Measuring range | Measuring current | Accuracy |
|---|---|---|---|
| 10Ω | 0.00mΩ~15.000Ω | 316mA | within ±0.02%±α±1d |
| 100Ω | 0.000Ω~150.00Ω | 1.00mA | within ±0.02%±1d |
| 1kΩ | 0.000Ω~1.5000kΩ | 316μA | |
| 10kΩ | 0.00Ω~150.00kΩ | 100μA | |
| 100kΩ | 0.0Ω~1500.0kΩ | 20μA | |
| 1MΩ | 0Ω~1.5000MΩ | 3.16μA | |
| 10MΩ | 0Ω~15.000MΩ | 0.63μA | within ±0.03%±α±1d |
| % | 1mΩ~15.000MΩ/±50.0% | refer to the above | refer to the above |
Compare with external standard thermistor/polymer PTC element of resistance value, Best suitable for percentage measurement
| Open-circuit voltage of measuring terminal | [Slow]:AC1~10cycles, [Fast]:0.1mS~99.9mS |
|---|---|
| Measuring method | 4-terminal measurement (Available to contact check) |
| Sampling time | [Remote start mode] about 1m [Fastest time] |
| Comparator set range | [Resistance measurement]:0~15,000 counts both for HI and LO |
| [% Measurement]:-99.99%~+50.00% both for HI and LO [Standard 100~10000:±50.00%] |
|
| Indication of comparator's comparison result | Comparator judgment result indication LO/LLG/LG/GO/HG/HHG/HI, judgment LED indication, buzzer setup |
| Machine interface control signal (Connector:57-40240 Equivalent) |
Remote start input:"L" [0V] →"H" [DC12V] start |
| Remote hold input:Open and "H" [DC12V] :Free run/"L" [0V]:Hold | |
| comparison output [7 common signal classification]: open collector:max.80V, 100mA |
|
| contact error output [CONT-E]:open collector:max.80V, 100mA | |
| end of comparison output [EOC]:open collector:max.80V, 100mA | |
| Operation condition | [Temp.] +5℃~+40℃ [Humidity] less than 85% (Disabled when condensation) |
| Power supply | AC100V/117V/220V/240V selectable, 50/60Hz, about 60VA |
| Outer dimension | 333 (W) ×99 (H) ×300 (D) mm (excluding protruding parts such as rubber legs, etc.) |
| Mass | about 4kg |
The Outline
AX-1137A/B indicates percentage deviations compare resistance value of thermistor polymer PTC element with external standard thermistor polymer PTC element.
Adding comparator of 2 layers (1137A) 5 layers (1137B) outputs to judge up to 7 classifications compare with the setup value.
Either GP-IB or centronics output can be equipped as an option.
Cần thêm thông tin hoặc tư vấn kỹ thuật?
Nếu Quý khách cần thêm thông tin chi tiết về sản phẩm, thông số kỹ thuật, ứng dụng thực tế hoặc yêu cầu mẫu thử, đội ngũ của chúng tôi luôn sẵn sàng hỗ trợ.
📞 Điện thoại: (024) 6687 2330
📧 Email: sale@pemax-mte.com
Hoặc gửi yêu cầu qua biểu mẫu liên hệ hoặc trao đổi trực tiếp qua hộp chat ở góc phải màn hình.