Easily to auto calibration, for anybody
?
● Auto calibration for digital resistance checker [AX-162D/AX-162E], one touch, high accuracy
● Unnecessary for standard resistance external attached
● Unnecessary for mastering calibration, experience, possible to calibrate easily and quick
● Possible to use as a standard resistance, high accuracy, high stability
● Possible to transfer printing out before calibration, after calibration
Specifications
The accuracy of installed standard resistance (at 23℃)
Reference resistance value | Accuracy | Rating | Temperature coefficient |
---|---|---|---|
SHORT | less than 1μΩ | within 0.5A | ---- |
10mΩ | less than ±0.02% | within ±10ppm/℃ | |
100mΩ | less than ±0.01% | ||
1Ω | less than ±0.005% | within ±5ppm/℃ | |
10Ω | within 1W | ||
100Ω | |||
1kΩ | within 0.5W | ||
10kΩ | within 50V | ||
100kΩ | |||
1MΩ | |||
10MΩ | |||
100MΩ | less than ±0.03% |
Time required for calibration | [(Tm×no1+Tm×no2+Tm×no3+Tm×nf+Tm×ns+600) ×10 Range] msec
|
|
---|---|---|
Operation condition | [Temp.] +10℃~+40℃, [Humidity] less than 70% | |
Power supply | AC 85V~265V, 50/60Hz, about 15VA | |
Outer dimension | 333 (W) × 99 (H) × 300 (D) mm (excluding protruding parts such as rubber legs, etc.) |
|
Weight | about 3.5kg |
The Outline
AX-16000A is auto calibration to calibrate measuring instrument AX-162D, high-speed measurement major unit of tip resistor, totally automatic (one touch).
It can be printed out the date before calibration, and after calibration connecting to our product AX-80B printer, also can be brought to realization to simplify keeping a date for periodical calibration.