Personal LCR Meter with comparator
Judgment indications [GO] [NG] built-in digital comparator
● Measuring range: L・・・0.1μH~199.9H
C・・・0.1pF~1999μF
R・・・1mΩ~1999kΩ
D・・・0.001~1.999
● L,C,R 3½, D 3½ digit, LED display
● Sampling time:10 times/sec. approx.
(Auto range changeover time:1 step 100 msec.)
● L,C,R and D, analog voltage output
(full scale DC2V)
Specifications
※The following accuracy is representative value
Measurement unit | |||
---|---|---|---|
Measuring signal and Accuracy (23℃±5℃) |
L | 0.1μH~199.9H | ±(0.5% of rdg+2digit+0.2μH) TYP |
C | 0.1pF~1999μF | ±(0.3% of rdg+2digit+0.2pF) TYP | |
R | 1mΩ~1999kΩ | ±(0.3% of rdg+2digit+2mΩ) TYP | |
D | 0.001~1.999 | ±(0.5% of rdg+2digit) TYP | |
Measuring frequency | 1kHz±5% |
Measuring signal level | L | 10mA~1μA | Differs according to the range. |
---|---|---|---|
C | 1V~0.01V | ||
R | 100mA~1μA | ||
Range changeover | Auto or manual | ||
Display | L,C,R 3½ digit, LED display | ||
Measuring time | Less than one second to stabilize the measured value | ||
Sampling time | 10 times per second approx. (Auto range changeover time:1 step 100 milli second) |
||
Changeover of measuring item | L and D, C and D, R push-button selectable | ||
Measuring mode | L | Series equivalent circuit | |
C | 200pF, 2nF, 20nF, 200nF range (parallel equivalent circuit) 2μF, 20μF, 200μF, 2000μF range (series equivalent circuit) |
||
R | Resistance, parallel resistance in inductance measurement | ||
Measuring terminal | 5 terminals | ||
Output signal | L, C, R and D, analog voltage output (full scale DC2V) | ||
External bias | 0~50V DC, 0.1A max. (capacitance measurement) | ||
Power supply | AC100V/117V/220V/240V selectable, 50/60Hz, about 15VA | ||
Outer dimension | about 333 (W) ×85 (H) ×245 (D) mm (excluding protruding parts such as rubber legs, etc.) |
||
Weight | about 3kg |
Comparator unit | |||
---|---|---|---|
Digital comparison mehod LCR, and D | |||
Judgment error | None | ||
Judgment time | about 1ms (Relay operate time) | ||
Setting range | 1999, D1999 LCR,-NG, with + NG | ||
Display GO / NG | LCR | -NG | Red LED |
+NG | |||
D | NG | ||
TOTAL | GO | Green LED | |
Buzzer | NG at buzzer, Buzzer GO, No buzzer can be switched |
||
Hold | Possible | ||
Relay output | LCR | -NG | One contact each transfer DC 30V 20W AC110V 33VA (Resistance load) |
+NG | |||
D | NG | ||
TOTAL | GO | ||
BCD Parallel Data out (Option) |
LCR | 1999 3½ digit and blank |
Parallel output fan out 1 |
D | 1999 3½ digit and D blank |
||
Measurement completes pulse width about 2ms positive pulse |
The Outline
Model AX-222, Digital LCR meter, can easily measure not only electronic part such as coil, electrostatic condenser, and resistor but also contact resistance like switch, relay and internal resistance of a battery and junction capacitance of semiconductor, and every elements.
Built-in comparator, it can be informed a judgment result by LED indication and buzzer also be printed out.
With a function of auto range and auto mode, it can measure to select the most approximate range automatically for measurement thing of an unknown value.
It is selectable a range by hand, will become shorten the measuring time by range switching.
Display shows a value of LCR for 3½ digits, at the same time indicates a value of D (loss factor) 3½ digit when L or C measurement.
Analog voltage outputs proportional to the measurement value, it can be connected to analog recorder or analog comparator, and others.